This website uses cookies primarily for visitor analytics. Certain pages will ask you to fill in contact details to receive additional information. On these pages you have the option of having the site log your details for future visits. Indicating you want the site to remember your details will place a cookie on your device. To view our full cookie policy, please click here. You can also view it at any time by going to our Contact Us page.

BROWSE PRODUCTS
 

High-resolution micro-ohmmeter has increased accuracy

TTi (Thurlby Thandar Instruments) has improved its BS407 micro-ohmmeter by increasing its accuracy at the lowest values of resistance.

High-resolution micro-ohmmeter has increased accuracy

The BS407 can resolve resistance down to 1 microhm, and has ranges with full scales from 2 milliohms up to 20 kilohms with a basic accuracy of 0.1%. It uses a 4-terminal measurement system via high quality Kelvin clip leads, and has a maximum excitation current of 250 mA. Current reversal can be selected to identify errors from thermal EMF.

A front-panel operated 'clamp' switch limits the maximum voltage across the unknown to 20 mV. This is a requirement for the measurement of switch contact resistances to international standards. Further front-panel switches allow for current reversal and for current diversion. The latter enables in-situ zero adjustments to be made.

The BS407 operates from built-in NiMh rechargeable batteries. Battery operation eliminates the noise that can be created by ground loops on AC line powered equipment. The instrument is fully portable, and has its own battery charger built in.